Sonicator 740 Therapeutic Ultrasound
*This item must be ordered by a licensed healthcare professional.
- Microprocessor-controlled, therapeutic ultrasound unit with enhanced reliability and ease of use
- Large, back-lit liquid crystal display shows treatment settings and status
- Large controls provide one touch entry for treatment parameters
- May be delivered in continuous, or 10%, 20% or 50% pulsed modes
- Comes with a dual-frequency 5-cm applicator, with direct crystal-to-patient contact
- Applicator is contoured for easy ultrasound application and has a coupling LED in the handle that lights up when adequate coupling is achieved
- Applicator attaches to the Sonicator 740 using the universal applicator cable
Sonicator 740 Specifications
General Specifications:
ETL and C-ETL Listed: Model ME 740 (9801427)
Input: 100-240 VAC +/-10%
Frequency: 60Hz +/-3 and 50Hz +/-3
Input Current: less than 0.6 amps
Pack rated at 10.8Vand 4.8Ah
Patents: U.S. and foreign patents applied for and granted including U.S. Patent
Numbers 4,966,131 and 5,095,890
Timer Accuracy: +/-10% of the indicated time
Maximum treatment time: 29 minutes
Size: 13.5 in (L) x 9.5 in (W) x 7 in (H)
Weight: 3.75
Ultrasonic Generator Specifications:
Frequency: 1.0 MHz +/-10%
3.2 MHz +/-10%
3.3 MHz +/-10%
Modes: Continuous
Pulsed 10, 20 and 50% Duty cycle
Pulse repetition rate: 100 Hz +/-5% (Pulse Mode)
Pulse duration: 1, 2 or 5 msec +/-5%
Temporal peak/average
Intensity ratio: 10:1, 5:1 and 2:1 +/-5%
Maximum output power: 22 W with a 10 cm2 applicator, (ME 7410)
11 W with 5 cm2 applicator, (ME 7413)
2.2 W with a 1 cm2 applicator (ME 7431)
Maximum intensity: 2.2 W/cm2
Indication accuracy: +/-20% (for any level above 10% of maximum)
Ultrasonic Applicator Specifications:
Piezoelectric discs: The output transducer utilizes a barium titanate disc with a specially
coated face.
Individual Applicator Specifications:
Applicator Part Number Frequency ERA
ME 7513 1 and 3.2 MHz +/-10% 5 cm2 +/-20%
Maximum Beam
Uniformity Ratio: 6:1